The Type SACA- auxiliary
air fume hoods have been engineered to reduce the amount of
costly conditioned air being exhausted from the room.
The large volume of air drawn from a room by a conventional
type fume hood places a great burden on the rooms air conditioning
and heating systems and greatly reduces the efficiency, while
increasing the operational cost, of these systems.
With the SACA- auxiliary
air fume hood, up to 70% of the air being exhausted by the
hood is supplementary air - air introduced from an outside
source at the top of the hood and distributed into the hood
around the entire outside face of the hod (top, both sides
and across the work surface). This supplementary air mingles
with about 30% of the conditioned room and sweeps the fumes
and vapors to the rear exhaust slots within the hood.
The supplementary (auxiliary) air is discharged into the hood
around the entire outside face of the hood (top, down both sides
and across the work surface) at a very low velocity, thereby
avoiding high velocity air streams blasting at the front of the
hood and into the chamber (which could be very uncomfortable
to the hood operator).
The type SACA- auxiliary
air fume hood is carefully engineered with the additional feature
of an air foil surface which is curved to the air flow path,
creating a smoother air flow pattern which can remove air effectively
at a lower face velocity. Thus, less room air is used than
a standard or conventional fume hood.
The SACA- fume
hood superstructures are furnished with:
- The hood exterior is fabricated
of 18 gauge, zinc-coated and bondarized steel with a chemical
resistant enamel to match the base cabinet (furnished upon
request).
- The interior is made of 1/4" thick,
special heat, solvent and reagent resistant composition stone
that is treated with a chemical resistant, clear vinyl for
added protection. (Additional hood liners are available upon
request, including a white modified resin/fiberglass, fiberglass
reinforced polyester (FRP), and white polypropylene.)
- The fume hood baffles have two
adjustable air flow dampers at the top and bottom for adjustment
to light or heavy gases.
- The interior of the hood is illuminated
with a vapor-proof incandescent light fixture for added visibility.
- A vertically rising front sash
has a 1/4" thick, laminated safety glass window that
is counter balanced with lead weights for fingertip movement.
Sash stops are available, upon request, to further restrict
the face opening of the hood.
Typical SACA- Fume
Hoods
| Catalog
#r |
Hood
Length |
Face
Opening w/ sash raised |
Exhaust
CFM @ 100 FPM face velocity |
Exh
S.P. |
Aux.
Air CFM-70% of Exhaust |
Aux.
Air S.P. |
Standard
Exhaust Blower Included |
Standard
Blowers
Included
Exhaust
Supply |
| SACA-4B |
48" |
7.5 SF |
750 CFM |
1/4" |
750 CFM |
1/2" |
D8C |
KE-9 |
| SACA-5B |
60" |
10.0
SF |
1000
CFM |
1/4" |
1000
CFM |
1/2" |
D9C |
D8C |
| SACA-6B |
72" |
12.5
SF |
1250
CFM |
3/8" |
1250
CFM |
1/2" |
KE-11 |
D8C |
| SACA-7B |
84" |
15.0
SF |
1500
CFM |
1/4" |
1500
CFM |
5/8" |
KE-12 |
KE-11 |
| SACA-8B |
96" |
17.5
SF |
1750
CFM |
1/4" |
1750
CFM |
5/8" |
KE-12 |
KE-11 |
| SACA-4 |
48" |
7.5 SF |
750 CFM |
1/4" |
750 CFM |
1/2" |
None |
None |
| SACA-5 |
60" |
10.0
SF |
1000
CFM |
1/4" |
1000
CFM |
1/2" |
None |
None |
| SACA-6 |
72" |
12.5
SF |
1250
CFM |
3/8" |
1250
CFM |
1/2" |
None |
None |
| SACA-7 |
84" |
15.0
SF |
1500
CFM |
1/4" |
1500
CFM |
5/8" |
None |
None |
| SACA-8 |
96" |
17.5
SF |
1750
CFM |
1/4" |
1750
CFM |
5/8" |
None |
None |
| *Note: |
Overall height shall be 9-0" (108")
when installed on to a 35-3/4" high base cabinet.
Fume hoods
may also be furnished without the support cabinet(s) and/or
work surfaces. |
| Note: |
A complete system, including chemical resistant duct
work, is available. |
Optional Features (but not limited to):
Work
Surface
- Cast epoxy resin, flat
or dished.
- "Solid Bond" Stainless
Steel
|
Service
Fixtures
- R400V remote control valve assembly with angled serrated
hose connection. Specify Index.
- R410V remote control water faucet with swivel goose
and hose connection for use with either cold or hot water.
Specify Index.
- R410V-MIX remote control mixing faucet with swivel
gooseneck and hose connection for use with hot and cold
water.
- A variety of deck mounted faucets and turret assemblies.
- Factory prepipe the service fixtures to either the
upper or lower rear of the hood.
|
Sinks,
Cup Sinks
- Epoxy resin
- Polypropylene
- Stainless steel
|
Electrical
Items
- Duplex or single receptacles. Specify voltage and amperage
required.
- Motor starter switch or on/off switch for exhaust
blower. Specify if single or three phase and voltage.
|
Light
Fixtures
- Two tube fluorescent light fixture (less bulbs).
- Explosion-proof, 150-300 watt incandescent light fixture
(less bulb).
|
Base
Cabinets
- Standard double cupboard cabinet
- Acid storage cabinet with composition stone lining.
(Vented doors and/or vent assembly to the hoods
work surface are optional.)
- Solvent storage cabinet. (Vent assembly to the hoods
work surface is optional.)
|
Safety
Guard Low Face Velocity Alarm
- Complies with OSHA regulations
- Factory installed and wired
|
CLICK
HERE TO VIEW SPECS 1 2 |